The Malaysian Journal of Analytical Sciences Vol 12 No1 (2008): 132 141
SQUARE WAVE CATHODIC
STRIPPING VOLTAMMETRIC TECHNIQUE
FOR DETERMINATION OF AFLATOXIN B1 IN GROUND
NUT SAMPLE
Mohamad Hadzri Yaacob1, Abdull Rahim Hj. Mohd. Yusoff
2
and Rahmalan Ahamad 2
1School
of Health Sciences, USM, 16150 Kubang Krian, Kelantan, Malaysia
2Chemistry Dept., Faculty of Sciences, UTM, 81310 Skudai,
Johor, Malaysia
Abstract
An
electroanalytical method has been developed for the detection and determination
of 2,3,6a,9a-tetrahydro-4-methoxycyclo penta[c] furo[3,2:4,5] furo [2,3-h][l]
benzopyran-1,11-dione (aflatoxin B1, AFB1) by a square wave cathodic stripping voltammetric (SWSV)
technique on a hanging mercury drop electrode (HMDE) in aqueous solution with
Britton-Robinson Buffer (BRB) at pH 9.0 as the supporting electrolyte. Effect
of instrumental parameters such as accumulation potential (Eacc), accumulation
time (tacc), scan rate (v), square wave frequency, step potential and pulse
amplitude were examined. The best condition were found to be Eacc of -0.8 V, tacc of
100 s, v of 3750 mVs-1, frequency of 125 Hz, voltage step of 30 mV and
pulse amplitude of 50 mV. Calibration curve was linear in the range of 0.01 to
0.15 uM with a detection limit of 0.125 x 10-8 M. Relative standard deviation
for a replicate measurement of AFB1 (n = 5) with a concentration of 0.01 ΅M was
0.83% with a peak potential of -1.30 V (against Ag/AgCl). The recovery values obtained in spiked ground
nut elute sample were 94.00 ± 0.67 % for 3.0 ppb, 91.22 ± 1.56 % for 9 ppb and
92.56 ± 2.00 % for 15.0 ppb of AFB1. The method was applied for the
determination of the AFB1 in ground nut samples after extraction and clean-up
steps. The results were compared with that obtained by high performance liquid
chromatography (HPLC) technique.
Keywords: square wave stripping voltammetry,
HMDE, aflatoxin B1, ground nut.
References
1. Akiyama, H. Goda, Y. Tanaka, T and Toyoda, M. 2001. Determination of aflatoxins B1, B2, G1 and G2 in spices
using a multifunctional
column clean-up. J. Chromatogr.
A. 932: 153-157.
2. Yoruglu, T, Oruc, H.H. and
Tayar, M. 2005. Aflatoxin M1 levels in cheese samples
from some provinces
of Turkey. Food Control
16(10): 883-895.
3. World
Health Organisation in: IARC Monograph on the evaluation of carcinogenic risk to human, WHO, Lyon, 1987.
4. Hall,
A.J., Wild, C.P in:
Eaton, D.L., Groopman, J.D. (Eds) 1994. The toxicology of aflatoxins: human health, veterinary and agricultural significance. Academic, New York.
5. Malaysian Food Act 1983. Act No 281, Reviewed 2002, Food
Quality Control Dept, Ministry of Health, Malaysia.
6. Bicking,M.K.L., Kniseley,
R.N
and Svec, H.J. 1983. Coupled-Column for quantitating low levels of aflatoxins. J . Assoc. Off.
Anal. Chem. 66: 905-908.
7. Beebe,
B.M. 1978.
Reverse phase high pressure liquid chromatographic determination of aflatoxins in foods. J.
Assoc. Off. Anal. Chem. 81: 1347-1352.
8. Cepeda, A., Franco, C.M., Fente, C.A., Vazquez, B.I., Rodriguez, J.L., Prognon, P. and Mahuzier, G. 1996.
Postcolumn excitation of aflatoxins using cyclodextrins in liquid chromatography for food analysis. J. Off.
Chrom. A 721: 69-74.
9. Garner, R.C., Whattam, M.M.,
Taylor, P.J.L. and Stow, M.W 1993. Analysis of United Kingdom
purchased spices for aflatoxins using immunoaffinity column clean up procedure followed up by high-performance liquid chromatographic analysis and post-column derivatization with pyridium
bromide perbromide
J. Chromatogr 648: 485-490
10. Beaver, R.W., James, M.A. and Lin ,T.Y. 1991. ELISA-based screening test with liquid chromatography for the determination of aflatoxin in corn.
J. Assoc.Off. Anal. Chem ,
74: 827 829
11. Garden,
S.R. and Strachan, N.J.C. 2001. Novel colorimetric immunoassay for the detection
of aflatoxin B1 Anal. Chim. Acta 444:
187-191.
12. Braitina, Kh. Z., Malakhova, N.A. and Stojko, Y. 2000.
Stripping voltammetry in environmental
and food analysis Fresenius J Anal. Chem.
368: 307-325
13. Skrzypek, S.,
Ciesielski, W., Sokolowski, A., Yilmaz, S.
and Kazmierczak, D. 2005. Square
wave adsorption stripping voltammetric determination of famotidine in urine. Talanta, 66(5): 1146-1151
14. Economou,
A., Bolis, S.D., Efstathiou, C.E. and Volikakis, G.J 2002. A virtual
electrochemical instrument for square wave voltammetry.Anal. Chim. Act. 467: 179-188.
15. Yaacob, M.H., Mohd Yusoff, A.R.
and Ahmad, R. 2003. Cyclic voltammetry of AFB2 at the mercury electrode. Simposium
Kimia Malaysia ke 13 (SKAM-13), 9
11th September 2003, Kucing, Sarawak,
Malaysia.
16. Standard
procedure for determination of aflatoxins 2000. Chemistry Dept. Penang Branch, Ministry of Science,
Technology and Innovation (MOSTI), Malaysia unpublished report.
17. Yaacob, M.H., Mohd Yusoff, A.R. and
Ahamad, R. 2005. Determination of the aflatoxin B1 in ground nut by
differential pulse cathodic stripping voltammetry technique. 2nd
National Seminar On Chemistry, 14th
April 2005, Universiti Sumatera Utara, Medan, Indonesia.